What is Electron
Microscopy?
The electron microscope is a type of microscope that uses a beam of
electrons to create an image of the specimen. It is capable of much higher
magnifications and has a greater resolving power than a light microscope, allowing
it to see much smaller objects in finer detail. They are large, expensive
pieces of equipment, generally standing alone in a small, specially designed
room and requiring trained personnel to operate them.
The History of EM
By the middle of the 19th century, microscopists had accepted that it was
simply not possible to resolve structures of less than half a micrometre with a
light microscope because of the Abbe’s formula, but the development of the
cathode ray tube was literally about to change the way they looked at things;
by using electrons instead of light!
Hertz (1857-94) suggested that cathode rays were a form of wave motion
and Weichert, in 1899, found that these rays could be concentrated into a small
spot by the use of an axial magnetic field produced by a long solenoid. But it was not until 1926 that Busch showed
theoretically that a short solenoid converges a beam of electrons in the same
way that glass can converge the light of the sun, that a direct comparison was
made between light and electron beams. Busch should probably therefore be known
as the father of electron optics.
In 1931 the German engineers Ernst Ruska and Maximillion Knoll succeeded
in magnifying and electron image. This
was, in retrospect, the moment of the invention of the electron microscope but
the first prototype was actually built by Ruska in 1933 and was capable of
resolving to 50 nm. Although it was
primitive and not really fit for practical use, Ruska was recognised some 50
years later by the award of a Nobel Prize. The first commercially available
electron microscope was built in England by Metropolitan Vickers for Imperial
College, London, and was called the EM1, though it never surpassed the
resolution of a good optical microscope.
The early electron microscopes did not excite the optical microscopists
because the electron beam, which had a very high current density, was
concentrated into a very small area and was very hot and therefore charred any
non-metallic specimens that were examined.
When it was found that you could successfully examine biological
specimens in the electron microscope after treating them with osmium and
cutting very thin slices of the sample, the electron microscope began to appear
as a viable proposition. At the
University of Toronto, in 1938, Eli Franklin Burton and students Cecil Hall,
James Hillier and Albert Prebus constructed the first electron microscope in
the New World. This was an effective,
high-resolution instrument, the design of which eventually led to what was to
become known as the RCA (Radio Corporation of America) range of very successful
microscopes.
Unfortunately, the outbreak of the Second World War in 1939 held back
their further development somewhat, but within 20 years of the end of the war
routine commercial electron microscopes were capable of 1 nm resolution.
Types of Electron Microscopes
All electron microscopes use electromagnetic and/or electrostatic lenses
to control the path of electrons. Glass
lenses, used in light microscopes, have no effect on the electron beam. The basic design of an electromagnetic lens
is a solenoid (a coil of wire around the outside of a tube) through which one
can pass a current, thereby inducing an electromagnetic field. The electron
beam passes through the centre of such solenoids on its way down the column of
the electron microscope towards the sample. Electrons are very sensitive to
magnetic fields and can therefore be controlled by changing the current through
the lenses.
The faster the electrons travel, the shorter their wavelength. The resolving power of a microscope is
directly related to the wavelength of the irradiation used to form an
image. Reducing wavelength increases
resolution. Therefore, the resolution of
the microscope is increased if the accelerating voltage of the electron beam is
increased. The accelerating voltage of the beam is quoted in kilovolts (kV). It
is now possible to purchase a 1,000kV electron microscope, though this is not
commonly found.
Although modern electron microscopes can magnify objects up to about two
million times, they are still based upon Ruska's prototype and the correlation
between wavelength and resolution. The electron microscope is an integral part
of many laboratories such as The John Innes Centre. Researchers can use it to examine
biological materials (such as microorganisms and cells), a variety of large
molecules, medical biopsy samples, metals and crystalline structures, and the
characteristics of various surfaces.
Nowadays, electron microscopes have many other uses outside
research. They can be used as part of a
production line, such as in the fabrication of silicon chips, or within
forensics laboratories for looking at samples such as gunshot residues. In the arena of fault diagnosis and quality
control, they can be used to look for stress lines in engine parts or simply to
check the ratio of air to solids in ice cream!
Transmission Electron Microscope (TEM)
The original form of electron microscopy, Transmission electron microscopy
(TEM) involves a high voltage electron beam emitted by a cathode and formed by
magnetic lenses. The electron beam that has been partially transmitted through
the very thin (and so semitransparent for electrons) specimen carries
information about the structure of the specimen. The spatial variation in this
information (the "image") is then magnified by a series of magnetic
lenses until it is recorded by hitting a fluorescent screen, photographic
plate, or light sensitive sensor such as a CCD (charge-coupled device) camera.
The image detected by the CCD may be displayed in real time on a monitor or
computer.
Transmission electron microscopes produce two-dimensional, black and white
images.
Resolution of the TEM is also limited by spherical and chromatic
aberration, but a new generation of aberration correctors has been able to
overcome or limit these aberrations. Software correction of spherical
aberration has allowed the production of images with sufficient resolution to
show carbon atoms in diamond separated by only 0.089 nm and atoms in silicon at
0.078 nm at magnifications of 50 million times. The ability to determine the
positions of atoms within materials has made the TEM an indispensable tool for
nano-technologies research and development in many fields, including
heterogeneous catalysis and the development of semiconductor devices for
electronics and photonics. In the life
sciences, it is still mainly the specimen preparation which limits the
resolution of what we can see in the electron microscope, rather than the
microscope itself.
At JIC we have a high voltage (200kV) TEM, which was installed in
2008. We have two digital cameras on it,
one is higher resolution than the other, so that the need for developing and
printing film has been negated. Our TEM
is designed for use with biological samples and is capable of resolving to
better than 1nm. It is also capable of
3-D tomography which involves taking a succession of images whilst tilting the
specimens through increasing angles, which can then be combined to form a
three-dimensional image of the specimen.
Scanning Electron Microscope (SEM)
Unlike the TEM, where the electrons in the primary beam are transmitted
through the sample, the Scanning Electron Microscope (SEM) produces images by
detecting secondary electrons which are emitted from the surface due to
excitation by the primary electron beam. In the SEM, the electron beam is
scanned across the surface of the sample in a raster pattern, with detectors
building up an image by mapping the detected signals with beam position.
SEM image of a fly's foot plate
showing the drawing of a fly's foot
SEM image of a fly's foot taken at JIC in 2006 From "Micrographia", by Robert Hooke, 1665: plate
showing the drawing of a fly's foot
TEM resolution is about an order of magnitude better than the SEM
resolution. Our TEM can easily resolve
details of 0.2nm. Our two SEMs at JIC
are both relatively recent acquisitions and are high-resolution instruments
capable of about 2 nm resolution on biological samples. Because the SEM image relies on electron
interactions at the surface rather than transmission it is able to image bulk
samples and has a much greater depth of view, and so can produce images that
are a good representation of the 3D structure of the sample. SEM images are therefore considered to
provide us with 3D, topographical information about the sample surface but will
still always be only in black and white.
In the SEM, we use much lower accelerating voltages to prevent beam
penetration into the sample since what we require is generation of the
secondary electrons from the true surface structure of a sample. Therefore, it is common to use low KV, in the
range 1-5kV for biological samples, even though our SEMs are capable of up to
30 kV.
At JIC we currently have two SEMs, both with high-resolution capabilities,
digital imaging facilities and cryo-systems which enable them to be used for
looking at frozen-hydrated specimens.
Sample Preparation
Materials to be viewed in an electron microscope generally require
processing to produce a suitable sample. This is mainly because the whole of
the inside of an electron microscope is under high vacuum in order to enable
the electron beam to travel in straight lines.
The technique required varies depending on the specimen, the analysis required
and the type of microscope:
Cryofixation - freezing a specimen rapidly, typically to liquid nitrogen
temperatures or below, that the water forms ice. This preserves the specimen in
a snapshot of its solution state with the minimal of artefacts. An entire field
called cryo-electron microscopy has branched from this technique. With the
development of cryo-electron microscopy, it is now possible to observe
virtually any biological specimen close to its native state.
Fixation - a general term used to describe the process of preserving a
sample at a moment in time and to prevent further deterioration so that it
appears as close as possible to what it would be like in the living state,
although it is now dead. In chemical fixation for electron microscopy,
glutaraldehyde is often used to crosslink protein molecules and osmium
tetroxide to preserve lipids.
Dehydration - removing water from the samples. The water is generally replaced with organic
solvents such as ethanol or acetone as a stepping stone towards total drying
for SEM specimens or infiltration with resin and subsequent embedding for TEM
specimens.
Embedding - infiltration of the tissue with wax (for light microscopy) or
a resin (for electron microscopy) such as araldite or LR White, which can then
be polymerised into a hardened block for subsequent sectioning.
Sectioning - the production of thin slices of the specimen. For light microscopy, the sections can be a
few micrometres thick but for electron microscopy they must be very thin so
that they are semitransparent to electrons, typically around 90nm. These
ultra-thin sections for electron microscopy are cut on an ultramicrotome with a
glass or diamond knife. Glass knives can easily be made in the laboratory and
are much cheaper than diamond, but they blunt very quickly and therefore need
replacing frequently.
Staining - uses heavy metals such as lead and uranium to scatter imaging
electrons and thus give contrast between different structures, since many
(especially biological) materials are nearly "transparent" to the
electron beam. By staining the samples with heavy metals, we add electron
density to it which results in there being more interactions between the
electrons in the primary beam and those of the sample, which in turn provides
us with contrast in the resultant image.
In biology, specimens can be stained "en bloc" before
embedding and/or later, directly after sectioning, by brief exposure of the
sections to solutions of the heavy metal stains.
Freeze-fracture and freeze-etch - a preparation method particularly useful
for examining lipid membranes and their incorporated proteins in "face
on" view. The fresh tissue or cell suspension is frozen rapidly
(cryofixed), then fractured by simply breaking or by using a microtome while
maintained at liquid nitrogen temperature. The cold, fractured surface is
generally "etched" by increasing the temperature to about -95°C for a
few minutes to let some surface ice sublime to reveal microscopic details. For
the SEM, the sample is now ready for imaging.
For the TEM, it can then be rotary-shadowed with evaporated platinum at
low angle (typically about 6°) in a high vacuum evaporator. A second coat of
carbon, evaporated perpendicular to the average surface plane is generally
performed to improve stability of the replica coating. The specimen is returned
to room temperature and pressure, and then the extremely fragile
"shadowed" metal replica of the fracture surface is released from the
underlying biological material by careful chemical digestion with acids,
hypochlorite solution or SDS detergent. The floating replica is thoroughly
washed from residual chemicals, carefully picked up on an EM grid, dried then
viewed in the TEM.
Sputter Coating - an ultra-thin coating of electrically-conducting
material, deposited by low vacuum coating of the sample. This is done to
prevent charging of the specimen which would occur because of the accumulation
of static electric fields due to the electron irradiation required during
imaging. It also increases the amount of secondary electrons that can be
detected from the surface of the sample in the SEM and therefore increases the
signal to noise ratio. Such coatings include gold, gold/palladium, platinum,
chromium etc.
Disadvantages of Electron Microscopy
Electron microscopes are very expensive to buy and maintain. They are
dynamic rather than static in their operation: requiring extremely stable high
voltage supplies, extremely stable currents to each electromagnetic coil/lens,
continuously-pumped high/ultra-high vacuum systems and a cooling water supply
circulation through the lenses and pumps. As they are very sensitive to
vibration and external magnetic fields, microscopes aimed at achieving high
resolutions must be housed in buildings with special services.
A significant amount of training is required in order to operate an
electron microscope successfully and electron microscopy is considered a
specialised skill.
The samples have to be viewed in a vacuum, as the molecules that make up
air would scatter the electrons. This means that the samples need to be
specially prepared by sometimes lengthy and difficult techniques to withstand
the environment inside an electron microscope. Recent advances have allowed
some hydrated samples to be imaged using an environmental scanning electron
microscope, but the applications for this type of imaging are still limited.
Artefacts
It must be emphasised from the outset that every electron micrograph is,
in a sense, an artefact. Changes in the
ultra-structure are inevitable during all the steps of processing that samples
must undergo: material is extracted, dimensions are changed and molecular
rearrangement occurs. The best thing we
can do is to keep these changes to a minimum by understanding the processes
involved so that we make informed choices of the best preparative procedures to
use for each sample. Artefacts present
themselves in many ways: there could be loss of continuity in the membranes,
distortion or disorganisation of organelles, empty spaces in the cytoplasm of
cells or sharp bends or curves in filamentous structures that are usually
straight, such as microtubules and so on. With experience, microscopists learn
to recognise the difference between an artefact of preparation and true
structure, mainly by looking at the same or similar specimens prepared in the
same or a different way.
Scanning electron microscopes usually image conductive or semi-conductive
materials best. Non-conductive materials can be imaged, either by an
environmental scanning electron microscope or more usually by coating the
sample with a conductive layer of metal. A common preparation technique is to
coat the sample with a layer of conductive material, a few nanometers thick,
such as 10nm of gold, from a sputtering machine. This process does, however, have the
potential to disturb delicate samples and cover some detail. When using chemical fixation and dehydration
as part of the sample preparation, there is often much shrinkage and collapse
of delicate structures and so, especially for our interests at JIC in botanical
specimens which are highly hydrated, we tend to use the cryo-fixation technique
which is far less prone to artefacts.
For the TEM, samples are generally prepared by exposure to many nasty
chemicals, in order to give good ultra-structural detail which may result in
artefacts purely as a result of preparation. This gives the problem of
distinguishing artefacts from genuine structures within the specimen,
particularly in biological samples. Scientists maintain that the results from
various preparation techniques have been compared, and as there is no reason
that they should all produce similar artefacts, it is therefore reasonable to
believe that electron microscopy features correlate with living cells. In
addition, higher resolution work has been directly compared to results from
X-ray crystallography, providing independent confirmation of the validity of
this technique. Recent work performed on unfixed, vitrified (rapidly frozen,
without the use of any chemicals, to form ice without any crystallisation)
specimens has also been performed, further confirming the validity of this
technique. However, even cryo-fixation techniques are not without their own
artefacts of preparation and ice crystal damage, due to the fact that as water freezes
it expands, is a common problem when trying to image a large specimen (greater
than 200 µm) which cannot be frozen rapidly enough to vitrify the water
No comments:
Post a Comment